Spiedigitallibrary, Conference Proceedings of SPIE
Volume
11886
Pages / Article-Number
118864
Keywords
Zebrafish embryo, hard X-ray computed tomography, microCT, conventional X-ray source, histological phenotyping, ethanol and paraffin embedding
Abstract
Hard X-ray micro computed tomography can be used for three-dimensional histological phenotyping of zebrafish embryos down to 1 µm or below without the need for staining or physical slicing. Current advances in ze- brafish embryo imaging, however, mostly rely on synchrotron radiation sources or highly advanced laboratory sources, which despite their evident strengths with regard to their beam properties and the implementation of phase contrast imaging techniques, lack accessibility. Therefore, we evaluated the performance of a conventional SkyScan 1275 laboratory µCT scanner in absorption contrast mode for the visualization of anatomical features in ethanol- and paraffin-embedded zebrafish embryos. We compare our results to readily available synchrotron data where 35 anatomical structures were identified. Despite having a more than ten times larger voxel length, approximately two thirds of the features could also be determined with laboratory microtomography. This could allow to monitor morphological changes during development or disease progression on large sample numbers, enabling the performance of preclinical studies in a local laboratory.