Data Entry: Please note that the research database will be replaced by UNIverse by the end of October 2023. Please enter your data into the system https://universe-intern.unibas.ch. Thanks
An Optimal Control Approach for {HIFU} Self-Scanning Treatment Planning
Book title (Conference Proceedings)
Medical Image Computing and Computer-Assisted Intervention − MICCAI
Volume
2
Place of Conference
Quebec
Publisher
Springer
Place of Publication
Cham
Pages
532-539
ISSN/ISBN
978-3-319-66184-1 ; 978-3-319-66185-8
Abstract
In noninvasive abdominal tumor treatment, research has focused on canceling organ motion either by gating, breath holding or tracking of the target. This paper is based on the novel self-scanning method which combines the advantages of the gated and the tracking method. This approach leverages the respiratory organ motion by holding the focal spot of the high intensity focused ultrasound (HIFU) device static for a given time, while it passively scans the tumor due to respiratory motion. This enables to use a lower-cost HIFU device. We present a planning method for such a system that is based on optimal control theory which optimizes the scanning path and the sonication intensities
simultaneously. The method minimizes treatment time and ensures complete tumor ablation according to the thermal dose under free-breathing. To verify our method, we simulated a tumor in two dimensions. The achieved treatment time performs on par to the gold-standard tracking method. Moreover, we measured the temperature profile of the HIFU device in a tissue-mimicking phantom to verify our temperature model.
Series title
Lecture Notes in Computer Science book series (LNCS)