Data Entry: Please note that the research database will be replaced by UNIverse by the end of October 2023. Please enter your data into the system https://universe-intern.unibas.ch. Thanks
Direct Calibration of a Laser Ablation System in the Projective Voltage Space
Book title (Conference Proceedings)
Medical Image Computing and Computer-Assisted Intervention MICCAI
Volume
9349
Place of Conference
München
Publisher
Springer
Pages
274-281
ISSN/ISBN
978-3-319-24552-2 ; 978-3-319-24553-9
Abstract
Laser ablation is a widely adopted technique in many contemporary medical applications. However, it is new to use a laser to cut bone and perform general osteotomy surgical tasks with it. In this paper, we propose to apply the direct linear transformation algorithm to calibrate and integrate a laser deflecting tilting mirror into the affine transformation chain of a sophisticated surgical navigation system, involving next generation robots and optical tracking. Experiments were performed on synthetic input and real data. The evaluation showed a target registration error of 0.3mm ± 0.2mm in a working distance of 150 mm.