Data Entry: Please note that the research database will be replaced by UNIverse by the end of October 2023. Please enter your data into the system https://universe-intern.unibas.ch. Thanks

Login for users with Unibas email account...

Login for registered users without Unibas email account...

 
Measurement setup and protocol for characterizing and testing radio frequency personal exposure meters
JournalArticle (Originalarbeit in einer wissenschaftlichen Zeitschrift)
 
ID 1634899
Author(s) Lauer, Oliver; Neubauer, Georg; Röösli, Martin; Riederer, Markus; Frei, Patrizia; Mohler, Evelyn; Fröhlich, Jürg
Author(s) at UniBasel Röösli, Martin
Frei, Patrizia
Mohler, Evelyn
Year 2012
Title Measurement setup and protocol for characterizing and testing radio frequency personal exposure meters
Journal Bioelectromagnetics
Volume 33
Number 1
Pages / Article-Number 75-85
Keywords electromagnetic fields, personal exposure, measurement accuracy, epidemiology
Abstract Body-worn radiofrequency electromagnetic field (RF-EMF) personal exposure meters (PEMs) have been increasingly used for exposure assessment in epidemiological research. However, little research on the measurement accuracy of these devices is available. In this article a novel measurement setup and a measurement protocol are presented for characterizing and testing PEMs. The whole setup and procedure is tested using two EME SPY 120 devices. The performance of the PEM was analyzed for absolute measurements in an anechoic chamber. Modulated signals representing the different services as real signals generated by appropriate testers were used. Measurement results were evaluated with respect to a root mean square detector. We found that measurement accuracy depends strongly on the carrier frequency and also on the number of occupied time slots for Time Division Multiple Access (TDMA)-based services. Thus, correction factors can only be derived if the distribution of the network configuration over the measurement time for all measurement points is available. As a result of the simplicity of the measurement setup and the straightforward measurement protocol, the possibility of fast validation leads to a higher accuracy in the characterization and testing of PEMs. Bioelectromagnetics. (c) 2011 Wiley-Liss, Inc
Publisher Wiley-Liss
ISSN/ISBN 1521-186X
edoc-URL http://edoc.unibas.ch/dok/A6094278
Full Text on edoc No
Digital Object Identifier DOI 10.1002/bem.20687
PubMed ID http://www.ncbi.nlm.nih.gov/pubmed/21755521
ISI-Number WOS:000297550000007
Document type (ISI) Article
 
   

MCSS v5.8 PRO. 0.322 sec, queries - 0.000 sec ©Universität Basel  |  Impressum   |    