Non-invasive central venous pressure measurement by compression ultrasound--a step into real life
JournalArticle (Originalarbeit in einer wissenschaftlichen Zeitschrift)
 
ID 1194879
Author(s) Thalhammer, Christoph; Siegemund, Martin; Aschwanden, Markus; Gassmann, Mirjam; Baumann, Ulrich A; Jaeger, Kurt A; Imfeld, Stephan
Author(s) at UniBasel Siegemund, Martin
Jäger, Kurt A.
Thalhammer, Christoph
Year 2009
Title Non-invasive central venous pressure measurement by compression ultrasound--a step into real life
Journal Resuscitation
Volume 80
Number 10
Pages / Article-Number 1130-6
Keywords Central venous pressure, Compression ultrasound, High-end ultrasound, Portable ultrasound, Emergency room, Training
Abstract AIM OF THE STUDY: Information on central venous pressure (CVP) is helpful in clinical situations like cardiac failure and sepsis. Compression ultrasound (CU) of a forearm vein has been shown to be a reliable method for CVP measurement when assessed by experienced personal under study conditions. To test the hypothesis, that CU can be reliably used for CVP measurement after a training program and using a portable ultrasound system, we investigated feasibility and accuracy of CU, comparing the results of vascular experts vs. trainees as well as high-end vs. a portable ultrasound system. METHODS: CU with non-invasive CVP measurement (CVP(ni)) was performed by four investigators in 50 patients of an intensive care unit and compared with invasive CVP measurement (CVP(i)). RESULTS: Feasibility was between 88 and 92% in the different investigator groups. CVP(ni) measurement showed a significant linear correlation (r=0.58-0.68; p<0.001) with CVP(i) in all groups. Mean difference between CVP(i) and CVP(ni) was 2.4+/-3.1mmHg and similar between the investigators. No differences were observed between measurements done by vascular experts and trainees, as well as between high-end and portable ultrasound systems. Further we demonstrated, that CU is able to detect changes in CVP during the respiratory cycle. CONCLUSIONS: After a training program CU can be used by non-vascular clinician for reliable CVP measurement with good quality portable systems. Furthermore, respiratory changes in CVP are detectable by CU. These data suggest CU to be an efficient method for measuring CVP under real life conditions.
Publisher Elsevier Science Ireland
ISSN/ISBN 0300-9572
edoc-URL http://edoc.unibas.ch/dok/A6005076
Full Text on edoc No
Digital Object Identifier DOI 10.1016/j.resuscitation.2009.06.027
PubMed ID http://www.ncbi.nlm.nih.gov/pubmed/19632026
ISI-Number WOS:000271336400010
Document type (ISI) Article
 
   

MCSS v5.8 PRO. 0.365 sec, queries - 0.000 sec ©Universität Basel  |  Impressum   |    
15/05/2024