Data Entry: Please note that the research database will be replaced by UNIverse by the end of October 2023. Please enter your data into the system https://universe-intern.unibas.ch. Thanks
SPECT-CT imaging in degenerative joint disease of the foot and ankle
Journal
The journal of bone and joint surgery : JBJS
Volume
91
Number
9
Pages / Article-Number
1191-6
Abstract
The precise localisation of osteoarthritic changes is crucial for selective surgical treatment. Single photon-emission CT-CT (SPECT-CT) combines both morphological and biological information. We hypothesised that SPECT-CT increased the intra- and interobserver reliability to localise increased uptake compared with traditional evaluation of CT and bone scanning together. We evaluated 20 consecutive patients with pain of uncertain origin in the foot and ankle by radiography and SPECT-CT, available as fused SPECT-CT, and by separate bone scanning and CT. Five observers assessed the presence or absence of arthritis. The images were blinded and randomly ordered. They were evaluated twice at an interval of six weeks. Kappa and multirater kappa values were calculated. The mean intraobserver reliability for SPECT-CT was excellent (kappa = 0.86; 95% CI 0.81 to 0.88) and significantly higher than that for CT and bone scanning together. SPECT-CT had significantly higher interobserver agreement, especially when evaluating the naviculocuneiform and tarsometatarsal joints. SPECT-CT is useful in localising active arthritis especially in areas where the number and configuration of joints are complex.
Publisher
British Editorial Society of Bone and Joint Surgery